DOSE MAP BLOOD IRRADIATOR DOSIMETRY SYSTEM: CUSTOMIZED BLOOD EQUIVALENT PHANTOM AND GAFCHROMIC EBT-XD FILM

Authors

  • Adiela Saiful Fazad
  • Muhammad Amir Nasir
  • Noramaliza Mohd Noor Noor Universiti Putra Malaysia

DOI:

https://doi.org/10.32896/ajmedtech.v3n2.44-54

Keywords:

Blood irradiator, Cesium-137, Gafchromic EBT-XD film, dose-mapping, Blood equivalent phantom

Abstract

Blood irradiation is done to sterilize and cease donor T-lymphocytes functionality to avoid Transfusion Associated-Graft Versus Host Disease (TA-GVHD). Dose mapping is the primary means of ensuring that the irradiation process is being conducted correctly. The aim of this study is to measure the minimum, maximum and central absorbed gamma radiation dose delivered at the newly customized blood equivalent phantom using a Gafchromic EBT-XD film. A Gammacell 3000 Elan Blood Irradiator with Cesium-137 source was used. To obtain the dose calibration curves, the films were placed at the center of a phantom and irradiated with the dose range from 5 Gy to 35 Gy with one unirradiated film as a control. The dose calibration curve was plotted using red, green and blue channels. While, for the dose mapping measurement, the irradiation exposure of 9.03 minutes was used to deliver a central dose of 25 Gy to the film. The response of the film will be compared with the GC Dose Mapping Report 2022 which used a water equivalent medium as a phantom. For the calibration curve, the red channel of the film was utilized in this study as it had a higher signal than the other channel. The doses obtained at central, minimum and maximum using EBT-XD film with customized blood equivalent phantom were in agreement with that obtained from GC Dose Mapping Report 2022 to be within ±7.6%, ±8.51% and ±9.95% respectively. We concluded that the customized blood equivalent phantom together with EBT-XD film has a potential to map the dose in blood irradiator accurately. The doses obtained from EBT-XD film were in the range of doses needed to inhibit the proliferation of the T-lymphocytes with central, minimum and maximum doses were 26.9 Gy, 17.2 Gy and 35.7 Gy respectively.

References

Bates, I., & Owusu-ofori, S. (2020). Blood Transfusion. In System-oriented Disease, 229–234.

Fast, L. D. (2015). Preventing transfusion-Associated graft-versus-host disease : state of the art. International Journal of Clinical Transfusion Medicine, (2), 1–6.

Syafinaz, N, Mohd Roslee, A. Nur Amalina, & M. N. Noramaliza. (2020.) Dose Mapping of Gamma Irradiation Chamber (GIC). Asian Journal of Medical Technology (AJMedTech) 2(1): 49-57.

Gòes, E. G., Borges, J. C., Covas, D. T., Orellana, M. D., Palma, P. V. B., Morais, F. R., & Pelá, C. A. (2006). Quality Control of Blood Irradiation: Determination T Cells Radiosensitivity To Cobalt-60 Gamma Rays.

Saglam, S., Cakir, A., & Seyfettin, K. (2011). Blood Irradiation. In Modern Approaches To Quality Control (pp. 335–347). Retrieved from http://www.intechopen.com/books/modern-approaches-to-quality-control/blood-irradiation

Moroff, G., & Luban, N. L. . (1997). The Irradiation of Blood and Blood Components to Prevent Graft-Versus-Host Disease: Technical Issues and Guidelines. Transfusion Medicine Reviews, 11(1), 15–26

T. Fearon, V. R. Criss, and N. L. C. Luban, “Blood irradiator dosimetry with BANG polymer gels,” vol. 45, pp. 1658–1662, 2005.

Best Theratronics Dosimetry Laboratory. (2022). GC 3000 Dose Mapping Report, Best Theratronics Ltd., Canada, 1-3.

Ashland. Gafchromic™ EBT-XD films. https://www.ashland.com/industries/medical/radiotherapy-films/ebtxd

Dreindl, R., Georg, D., & Stock, M. (2014). Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films. Zeitschrift für Medizinische Physik, 24(2), 153-163.

Palmer, A. L., Dimitriadis, A., Nisbet, A., & Clark, C. H. (2015). Evaluation of Gafchromic EBT-XD film, with comparison to EBT3 film, and application in high dose radiotherapy verification. Physics in Medicine & Biology, 60(22), 8741.

Olivo, R. A., Silva, M. V. D., Garcia, F. B., Soares, S., Rodrigues, V., & Moraes-Souza, H. (2015). Evaluation of the effectiveness of packed red blood cell irradiation by a linear accelerator. Revista brasileira de hematologia e hemoterapia, 37, 153-159.

K.S.A.K. Bakar, N.M. Noor, M.F. Hassan, F. Idris, D.A. Bradley. (2023). Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: A preliminary study, Radiation Physics and Chemistry, 209, 111010

Published

2023-07-31

How to Cite

Saiful Fazad, A., Nasir, M. A., & Noor, N. M. . N. (2023). DOSE MAP BLOOD IRRADIATOR DOSIMETRY SYSTEM: CUSTOMIZED BLOOD EQUIVALENT PHANTOM AND GAFCHROMIC EBT-XD FILM. Asian Journal Of Medical Technology, 3(2), 44–54. https://doi.org/10.32896/ajmedtech.v3n2.44-54